The study of probe loading capacity of HPLC monolithic columns based on organic porous polymers

Authors

  • Ю. В. Патрушев Patrushev Yuri V. - Ph.D. (chemistry), researcher. Boreskov Institute of Catalysis SB RAS, Novosibirsk, E-mail: patrush@catalysis.ru
  • Ю. С. Юдина Yudina Yulia S. - Student of the Novosibirsk state university, Novosibirsk, E-mail: juliaj94@mail.ru
  • В. Н. Сидельников Sidelnikov Vladimir N. - prof., grand Ph.D (chemistry), senior researcher, head of the Analytical laboratory. Boreskov Institute of Catalysis SB RAS, Novosibirsk, E-mail: vlad@catalysis.ru

DOI:

https://doi.org/10.17308/sorpchrom.2017.17/351

Keywords:

ВЭЖХ, монолитные колонки, пористые полимеры, 1-винилимидазол, 4- винилпиридин.

Abstract

The loading properties of HPLC monolithic columns based on 1-vinylimidazole and 4-vinylpyridine
were studied. These properties were compared with commercial packed column (Diasfer-250-ST) of the
same dimension based on divinylbenzene-styrene polymer.
It was shown that for dried monolith sorbents the textural characteristics (specific surface area, pore
volume and dimensions) can’t be used as a sample loading indicator. This is connected with a porous system
modification in the organic monolith placed in a «good» solvent. Diasfer column packaging material is not
affected by the solvent.
Therefore, proper selection of the mobile phase composition for investigated monolithic columns allow
to work with the large samples about 100 µg. This is significantly exceeds the quantity of the sample size
to a Diasfer packed column (30-40 µg).